Fluke Calibration 1552a Stik Thermometer Fluke Calibration 1552a Stik Thermometer Accurate and repeatable to ±0.05°C over its full range, the 1552A “Stik” Thermometer is the “gold standard” of industrial temperature calibration. P000282 Temperature Calibrators 0 VND Quantity: 1 Cái


  •  
  • Fluke Calibration 1552a Stik Thermometer

  • Post on 23-09-2015 10:39:44 AM - 1064 Views
  • Price: Contact

  • Accurate and repeatable to ±0.05°C over its full range, the 1552A “Stik” Thermometer is the “gold standard” of industrial temperature calibration.


Quantity
  • Finally, a digital substitute for your mercury-in-glass thermometers! Accurate and repeatable to ±0.05°C over its full range, the 1552A “Stik” Thermometer is the new “gold standard” of industrial temperature calibration. Whether working outdoors in environments where potentially explosive gases may be present or on the floor of a processing plant, the intrinsically-safe, battery operated, portable reference thermometer is designed to go where you work.

    Features at a glance:

    • Accuracy of ±0.05°C (±0.09°F) over full range
    • Intrinsically safe (ATEX and IECEx compliant)
    • Two models to choose from (-50°C to 160°C or -80°C to 300°C)
    • User-configurable temperature Trend/Stability indicator
    • Display temperature in °C or °F
    • Optional data logging to internal memory
    • 300-hour battery life
    • Percent battery-life and low-battery indicator
    • NVLAP-accredited, NIST-traceable calibration included
Specifications
Temperature range -80°C to 300°C
(-112°F to 572°F)
Accuracy (1 year) ±0.05°C (±0.09°F)
Display units °C, °F
Sensor type 100 Ω wire-wound PRT
Probe temperature coefficient 0.00385 Ω/Ω/°C nominal
Sensor length ≤ 30 mm
Sensor position (from sheath tip) 3 mm (0.1 in)
Minimum immersion depth 12 cm (4.8 in)
Probe sheath material Stainless steel
Response time 4.8 mm diameter probe (3/16 in): 14 seconds
6.35 mm diameter probe (1/4 in): 21 seconds
Probe hysteresis ±0.01°C
Temperature resolution Selectable: 0.1, 0.01, 0.001 (default 0.01)
Sample rate Selectable: 0.5 seconds, 1 second, 2 seconds (default 1 seconds)
Operating temperature range readout -10°C to 50°C (14°F to 122°F)
Temperature resolution Selectable: 0.1, 0.01, 0.001 (default 0.01)
Sample rate Selectable: 0.5 seconds, 1 second, 2 seconds (default 1 seconds)
Operating temperature range readout -10°C to 50°C (14°F to 122°F)
Storage temperature -20°C to 60°C (-4°F to 140°F) 0% to 95% RH, non-condensing
Optional data logging1 Up to 10,000 time-stamped readings stored to internal memory
Logging intervals1 2 seconds, 5 seconds, 10 seconds, 30 seconds, or 60 seconds;
2 minutes, 5 minutes, 10 minutes, 30 minutes, or 60 minutes
Damping Moving average of the most recent 2 to 10 readings
(ON/OFF, 2, 5, 10)
Communications RS-232 stereo jack (access calibration parameters only)
AC power None
DC power 3-AAA Batteries, typical battery life of 300 hours without LCD backlight
EMC compliance EN61326:2006 Annex C; CISPR II Edition 5.0-2009; Class B
Enclosure rating IP50
Electronics dimensions (H x W x D) 114 x 57 x 25 mm
(4.5 x 2.25 x 1.0 in)
Weight 196 g (6.9 oz)
Calibration (included) NVLAP-accredited, NIST-traceable
Characterization ITS-90
1. See ordering information for optional data logging configurations
     

 