Fluke 754 Documenting Process Calibrator-HART Fluke 754 Documenting Process Calibrator-HART Combining HART communication in a documenting process calibrator, the 754 is the ideal tool for calibrating and troubleshooting HART instrumentation. P000264 Temperature Calibrators 0 VND Quantity: 1 Cái


  •  
  • Fluke 754 Documenting Process Calibrator-HART

  • Post on 22-09-2015 11:23:26 PM - 936 Views
  • Price: Contact

  • Combining HART communication in a documenting process calibrator, the 754 is the ideal tool for calibrating and troubleshooting HART instrumentation.


Quantity
  • You’re going to love what we’ve done with your calibrator.

    The 754 is everything you needed and wished for in the 744 and more! HART communication capability is combined to deliver an integrated communicating calibrator. This rugged, reliable tool is ideal for calibrating, maintaining, and troubleshooting HART and other instrumentation.

    The 754 does the work of several tools - sourcing, simulating and measuring pressure, temperature, and electrical signals in one rugged, hand-held device. You’ll fall in love with the new improved graphical screen, Li-Ion battery for longer life, USB port and new accessories to complete the package.

    For documentation, the 754 automates calibration procedures and captures your data. And, of course, it helps you meet rigorous standards like ISO 9000, FDA, EPA and OSHA regulations.

    HART Capabilities

    The 754 is designed to take on nearly all the day-to-day tasks you now perform with a separate communicator. In fact, it offers most of the communication capabilities of a HART communicator.

    To create a seamless/paperless calibration management system consider adding Fluke DPCTrack2 Calibration Management software for use with your Fluke 753 and 754 or even legacy Fluke 743 and 744 calibrators.

    The Fluke 754 Documenting Process Calibrator-HART: Work smarter. Work faster.

     

    Measurement Accuracy
    Voltage DC Range/Resolution 1 Year 2 Years
    100.000 mV 0.02% + 0.005 mV 0.03% + 0.005 mV
    3.00000 V 0.02% + 0.00005 V 0.03% + 0.00005 V
    30.0000 V 0.02% + 0.0005 V 0.03% + 0.0005 V
    300.00 V 0.05% + 0.05 V 0.07% + 0.05 V
    Voltage AC 3.000 V (40 Hz to 500 Hz) / 0.001 V 0.5% + 0.002 V 1.0% + 0.004 V
    30.00 V (40 Hz to 500 Hz) / 0.01 V 0.5% + 0.02 V 1.0% + 0.04 V
    300.0 V (40 Hz to 500 Hz) / 0.1 V 0.5% + 0.2 V 1.0% + 0.2 V
    Current DC 30.000 mA 0.01% + 5 uA 0.015% + 7 uA
    110.00 mA 0.01% + 20 uA 0.015% + 30 uA
    Resistance 10.000 Ω 0.05% + 50 mΩ 0.07% + 70 mΩ
    100.00 Ω 0.05% + 50 mΩ 0.07% + 70 mΩ
    1.0000 kΩ 0.05% + 500 mΩ 0.07% + 0.5 Ω
    10.000 kΩ 0.1% + 10 Ω 0.15% + 15 Ω
    Frequency 1.00 to 110.00 Hz / 0.01 Hz   0.05 Hz
    110.1 to 1100.0 Hz / 0.1 Hz   0.5 Hz
    1.101 to 11.000 kHz / 0.001 kHz   0.005 kHz
    11.01 to 50.00 kHz / 0.01 kHz   0.05 kHz
    Source Accuracy
        1 Year 2 Years
    Voltage DC 100.000 mV 0.01% + 0.005 mV 0.015% + 0.005 mV
    1.00000 V 0.01% + 0.00005 V 0.015% + 0.0005 V
    15.0000 V 0.01% + 0.0005 V 0.015% + 0.0005 V
    Current DC 22.000 mA (source) 0.01% + 0.003 mA 0.02% + 0.003 mA
    Current sink (simulate) 0.02% + 0.007 mA 0.04% + 0.007 mA
    Resistance 10.000 Ω 0.01% + 10 mΩ 0.015% + 15 mΩ
    100.00 Ω 0.01% + 20 mΩ 0.015% + 30 mΩ
    1.0000 kΩ 0.02% + 0.2 Ω 0.03% + 0.3 Ω
    10.000 kΩ 0.02% + 3 Ω 0.03% + 5 Ω
    Frequency 0.1 to 10.99 Hz   0.01 Hz
    0.01 to 10.99 Hz   0.01 Hz
    11.00 to 109.99 Hz   0.1 Hz
    110.0 to 1099.9 Hz   0.1 Hz
    1.100 to 21.999 kHz   0.002 kHz
    22.000 to 50.000 kHz   0.005 kHz
    Technical Data
    Data log functions Measure functions Voltage, current, resistance, frequency, temperature, pressure
    Reading rate 1, 2, 5, 10, 20, 30, or 60 readings/minute
    Maximum record length 8000 readings (7980 for 30 or 60 readings/minute)
    Ramp functions Source functions Voltage, current, resistance, frequency, temperature
    Rate 4 steps/second
    Trip detect Continuity or voltage (continuity detection not available when sourcing current)
    Loop power function Voltage Selectable, 26 V
    Accuracy 10%, 18 V minimum at 22 mA
    Maximum current 25 mA, short circuit protected
    Maximum input voltage 50 V DC
    Step functions Source functions Voltage, current, resistance, frequency, temperature
    Manual step Selectable step, change with arrow buttons
    Autostep Fully programmable for function, start delay, stepvalue, time per step, repeat
    Environmental Specifications
    Operating temperature -10°C to +50°C
    Storage temperature -20°C to +60°C
    Dust/water resistance Meets IP52, IEC 529
    Operating altitude 3000 m above mean sea level (9842 ft)
    Safety Specifications
    Agency approvals CAN/CSA C22.2 No 1010.1-92, ASNI/ISA S82.01-1994, UL3111, and EN610-1:1993
    Mechanical and General Specifications
    Size 136 x 245 x 63 mm (5.4 x 9.6 x 2.5 in)
    Weight 1.2 kg (2.7 lb)
    Batteries Internal Battery Pack Li-ion: 7.2V,4400mAh, 30 Wh
    Battery life >8 hours typical
    Battery replacement Replace without opening calibrator; no tools required
    Side port connections Pressure module connector
    USB Connector to interface to your PC
    Digital instrument (HART) connector
    Connection for optional battery charger/eliminator
    Data storage capacity 1 week of calibration procedures results
    90 day specifications The standard specification interval for the 750 Seriesare 1 and 2 years.
    Typical 90 day measurement and source accuracy can be estimated by dividing the one year "% of reading" or "% of output" specifications by 2.
    Floor specifications, expressed as "% of full scale" or "counts" or "ohms" remain constant.
    Temperature, Resistance Temperature Detectors
    Degrees or % of reading - Type (α) Range °C Measure °C1
      1 year 2 year
    100 Ω Pt (385) -200 to 100
    100 to 800
    0.07°C
    0.02% + 0.05°C
    0.14°C
    0.04% + 0.10°C
    200 Ω Pt (385) -200 to 100
    100 to 630
    0.07°C
    0.02% + 0.05°C
    0.14°C
    0.04% + 0.10°C
    500 Ω Pt (385) -200 to 100
    100 to 630
    0.07°C
    0.02% + 0.05°C
    0.14°C
    0.04% + 0.10°C
    1000 Ω Pt (385) -200 to 100
    100 to 630
    0.07°C
    0.02% + 0.05°C
    0.14°C
    0.04% + 0.10°C
    100 Ω Pt (3916) -200 to 100
    100 to 630
    0.07°C
    0.02% + 0.05°C
    0.14°C
    0.04% + 0.10°C
    100 Ω Pt (3926) -200 to 100
    100 to 630
    0.08°C
    0.02% + 0.06°C
    0.16°C
    0.04% + 0.12°C
    10 Ω Cu (427) -100 to 260 0.2°C 0.4°C
    120 Ω Ni (672) -80 to 260 0.1°C 0.2°C
    Source current Source °C Allowable current2
      1 year 2 year  
    1 mA 0.05°C
    0.0125% + 0.04°C
    0.10°C
    0.025% + 0.08°C
    0.1 mA to 10 mA
    500 μA 0.06°C
    0.017% + 0.05°C
    0.12°C
    0.034% + 0.10°C
    0.1 mA to 1 mA
    250 μA 0.06°C
    0.017% + 0.05°C
    0.12°C
    0.034% + 0.10°C
    0.1 mA to 1 mA
    150 μA 0.06 C
    0.017% + 0.05°C
    0.12 C
    0.034% + 0.10°C
    0.1 mA to 1 mA
    1 mA 0.05°C
    0.0125% + 0.04°C
    0.10°C
    0.025% + 0.08°C
    0.1 mA to 10 mA
    1 mA 0.05°C
    0.0125% + 0.04°C
    0.10°C
    0.025% + 0.08°C
    0.1 mA to 10 mA
    3 mA 0.2°C 0.4°C 0.1 mA to 10 mA
    1 mA 0.04°C 0.08°C 0.1 mA to 10 mA
    1. For two and three-wire RTD measurements, add 0.4°C to the specifications.
    2. Supports pulsed transmitters and PLCs with pulse times as short as 1 ms
    Temperature, Thermocouples
    Type Source °C Measure °C Source °C
        1 year 2 years 1 year 2 years
    E -250 to -200 1.3 2.0 0.6 0.9
    -200 to -100 0.5 0.8 0.3 0.4
    -100 to 600 0.3 0.4 0.3 0.4
    600 to 1000 0.4 0.6 0.2 0.3
    N -200 to -100 1.0 1.5 0.6 0.9
    -100 to 900 0.5 0.8 0.5 0.8
    900 to 1300 0.6 0.9 0.3 0.4
    J -210 to -100 0.6 0.9 0.3 0.4
    -100 to 800 0.3 0.4 0.2 0.3
    800 to 1200 0.5 0.8 0.3 0.3
    K -200 to -100 0.7 1.0 0.4 0.6
    -100 to 400 0.3 0.4 0.3 0.4
    400 to 1200 0.5 0.8 0.3 0.4
    1200 to 1372 0.7 1.0 0.3 0.4
    T -250 to -200 1.7 2.5 0.9 1.4
    -200 to 0 0.6 0.9 0.4 0.6
    0 to 400 0.3 0.4 0.3 0.4
    B 600 to 800 1.3 2.0 1.0 1.5
    800 to 1000 1.0 1.5 0.8 1.2
    1000 to 1820 0.9 1.3 0.8 1.2
    R -20 to 0 2.3 2.8 1.2 1.8
    0 to 100 1.5 2.2 1.1 1.7
    100 to 1767 1.0 1.5 0.9 1.4
    S -20 to 0 2.3 2.8 1.2 1.8
    0 to 200 1.5 2.1 1.1 1.7
    200 to 1400 0.9 1.4 0.9 1.4
    1400 to 1767 1.1 1.7 1.0 1.5
    C 0 to 800 0.6 0.9 0.6 0.9
    800 to 1200 0.8 1.2 0.7 1.0
    1200 to 1800 1.1 1.6 0.9 1.4
    1800 to 2316 2.0 3.0 1.3 2.0
    L -200 to -100 0.6 0.9 0.3 0.4
    -100 to 800 0.3 0.4 0.2 0.3
    800 to 900 0.5 0.8 0.2 0.3
    U -200 to 0 0.6 0.9 0.4 0.6
    0 to 600 0.3 0.4 0.3 0.4
    BP 0 to 1000 1.0 1.5 0.4 0.6
    1000 to 2000 1.6 2.4 0.6 0.9
    2000 to 2500 2.0 3.0 0.8 1.2
    XK -200 to 300 0.2 0.3 0.2 0.5
    300 to 800 0.4 0.6 0.3 0.6
     

 

  Reader Comments

You did not use the site, Click here to remain logged. Timeout: 60 second