Fluke 725Ex Intrinsically Safe Multifunction Process Calibrator Fluke 725Ex Intrinsically Safe Multifunction Process Calibrator The 725Ex Intrinsically Safe Multifunction Process Calibrator P000274 Temperature Calibrators 0 VND Quantity: 1 Cái


  •  

Quantity
  • The Simply Powerful Intrinsic Safe Calibration Tool

    The 725Ex is an intrinsically safe multifunction process calibrator for usage in explosion endangered areas. These calibration tools source and measure almost all process parameters.

    Measure and source mA, volts, temperature (RTDs and thermocouples), frequency, ohms, and pressure, using optional pressure modules.

    Do you need to calibrate transmitters? The 725Ex split display lets you view input and output values simultaneously. For valve and I/P tests, you can source mA while measuring pressure. The 725Ex has auto-stepping and auto-ramping for remote testing, plus 25% stepping for fast linearity tests.

    From the moment you pick it up, the Fluke 725Ex is ready to perform. Its simple controls, without menus, make operation easy. Its memory functions make set up fast. And its rugged design lets it work as hard as you do.

    Measurement Accuracy
    Voltage DC 30.000 V 0.02% + 2 counts (upper display)
    10.000 V 0.02% + 2 counts (lower display)
    90.00 mV 0.02% + 2 counts
    -10.00 mV to 75.00 mV 0.025% + 1 count (via TC connector)
    Current DC 24.000 mA 0.02% + 2 counts
    Resistance 0.0 to 400.0 Ω 0.1 Ω (4-wire), 0.15 Ω (2- and 3-wire)
    401 to 1500 Ω 0.5 Ω (4-wire), 1 Ω (2- and 3-wire)
    1500 to 3200 Ω 1 Ω (4-wire), 1.5 Ω (2- and 3-wire)
    Frequency 2.0 to 1000.0 CPM 0.05% + 1 count
    1.0 to 1100.0 Hz 0.05% + 1 count
    1.00 to 10.00 kHz 0.05% + 1 count
    Sensitivity 1 V peak-to-peak-minimum
    Pressure Accuracy from 0.025% of range using any of 8 intrinsic safe pressure modules (for detailed specifications refer to pressure modules in options and accessories). Modules available for differential, gauge, vacuum, absolute, dual and high pressure.
    Source Accuracy
    Voltage DC 100.00 mV 0.02% + 2 counts
    10.000 V 0.02% + 2 counts
    -10.00 mV to 75.00 mV 0.025% + 1 count (via TC connector)
    Current DC 24.000 mA (source) 0.02% + 2 counts
    24.000 mA (simulate) 0.02% + 2 counts
    Resistance 15.0 to 400.0 Ω 0.15 Ω (exc. current 0.15 to 0.5 mA), 0.1 Ω (exc. current 0.5 to 2 mA)
    401 to 1500 Ω 0.5 Ω (excitation current 0.05 to 0.8 mA)
    1500 to 3200 Ω 1 Ω (excitation current 0.05 to 0.4 mA)
    Frequency 2.0 to 1000.0 CPM 0.05%
    1.0 to 1100.0 Hz 0.05%
    1.00 to 10.00 kHz 0.25%
    Waveform 5 V p-p squarewave, -0.1 V offset
    RTDs and Thermocouples
    Measure accuracy NI-120 0.2°C
    PT-100 (385) 0.33°C
    PT-100 (393) 0.3°C
    PT-100 (JIS) 0.3°C
    PT-200 (385) 0.2°C
    PT-500 (385) 0.3°C
    PT-1000 (385) 0.2°C
    Resolution 0.1°C
    J 0.7°C
    K 0.8°C
    T 0.8°C
    E 0.7°C
    R 1.8°C
    S 1.5°C
    B 1.4°C
    L 0.7°C
    U 0.75°C
    N 0.9°C
    Resolution J, K, T, E, L, N, U: 0.1°C, 0.1°F B, R, S: 1°C, 1°F
    XK 0.6°C
    BP 1.2°C
    Source accuracy NI-120 0.2°C
    PT-100 (385) 0.33°C
    PT-100 (393) 0.3°C
    PT-100 (JIS) 0.3°C
    PT-200 (385) 0.2°C
    PT-500 (385) 0.3°C
    PT-1000 (385) 0.2°C
    Resolution 0.1°C
      Accuracy stated for 4-wire measurement.
    J 0.7°C
    K 0.8°C
    T 0.8°C
    E 0.7°C
    R 1.4°C
    S 1.5°C
    B 1.4°C
    L 0.7°C
    U 0.75°C
    N 0.9°C
    Resolution J, K, T, E, L, N, U: 0.1°C, B, R, S: 1°C
    XK 0.6°C
    BP 1.2°C
    Specifications
    Ramp functions Source functions Voltage, current, resistance, frequency, temperature
    Ramps Slow ramp, Fast ramp, 25% step-ramp
    Loop power function Voltage 12 V
    Accuracy 10%
    Maximum current 22 mA, short circuit protected
    Step functions Source functions Voltage, current, resistance, frequency, temperature
    Steps 25% of range, 100% of range
    Safety specifications
    Agency approvals ATEX II 1 G Ex ia IIB 171°C
    I.S. Class I, Division 1 Groups B-D
    Mechanical and General Specifications
    Size 130 x 236 x 61 mm (5.188 x 9.291 x 2.402 in)
    Weight 0.85 kg (1.874 lbs.)
    Batteries 4 AA alkaline batteries
    Warranty Three year
    Battery replacement Separate battery compartment, accessible without breaking calibration seal
    Side port connections Pressure module connector
     

 

  Reader Comments

You did not use the site, Click here to remain logged. Timeout: 60 second