Fluke 719Pro Electric Pressure Calibrator Fluke 719Pro Electric Pressure Calibrator Fluke 719Pro Electric Pressure Calibrator with integrated electric pump-up P000286 Pressure Calibrators 0 VND Quantity: 1 Cái


  •  

Quantity

The 719Pro includes a full functioning loop calibrator that sources, simulates and measures mA signals and more making it the ideal test tool for calibrating high accuracy transmitters, pressure switches and pressure gauges.

Get the ultimate in measurement flexibility with the large backlit screen, which displays three parameters at once:

  • Pressure measurement from internal or external sensor
  • Sourced/simulated or measured mA values
  • Temperature measured by optional RTD probe
 
  • Measurement
    Fluke-719Pro-30G Range -12 to 30 psi / -0.8 to 2 bar
    Resolution 0.001 psi / 0.0001 bar
    Fluke-719Pro-150G Range -12 to 150 psi / -0.8 to 10 bar
    Resolution 0.01 psi / 0.001 bar
    Fluke-719Pro-300G Range -12 to 300 psi / -0.8 to 20 bar
    Resolution 0.01 psi / 0.001 bar
    Accuracy ±0.025% of full scale, 6 months
    ±0.035% of full scale, 1 year
    Temperature effect (all ranges) Add ±0.002% F.S./°C for temps outside of 15°C to 35°C
    Dry air and non-corrosive gas only
    No effect on accuracy on all functions from 15°C to 35°C
    Function Range Resolution Accuracy (1 year)
    mA DC (simulation and measurement) 0-24 mA DC 0.001 mA ±0.015% of reading ±2 counts
    1000 ohm maximum load in mA source, 26 V DC maximum voltage in mA simulate
    Volts DC (measurement only) 30 V DC 0.001 V ±0.015% of reading ±2 counts
    No AC voltage measurement, do not exceed 30 V DC
    Temperature (measurement only, Pt100 RTD) -50 to 150°C 
    (-50°F to 302°F)
    0.01°C
    0.01°F
    Temperature ±0.1°C (0.2°F)
    ±0.25°C (±0.45°F) combined uncertainty when using 720 RTD probe (optional accessory)
    Loop power supply 24 V DC N/A 24 mA at 24 V
    Temperature effect (all functions) - - Add ±0.002% F.S./°C for temps outside of 15°C to 35°C
    No effect on accuracy on all functions from 15°C to 35°C
    Environmental
    Operating temperature -10°C to +50°C
    Storage temperature (with batteries) Per battery manufacturer's specification within storage specification below
    Storage temperature (without batteries) -20°C to +60°C
    Power Requirements
    Battery Eight (8) standard Alkaline AA batteries
    Battery life 300 pump cycles to 150 psi minimum; 1000 pump cycles to 30 psi; 100 pump cycles to 300 psi minimum
    Physical
    Dimensions (H x W x D) 24.13 x 11.18 x 7 cm (9.5 x 4.4 x 2.8 in)
    Weight With holster 1.2 kg (2.5 lb)
    Connections
    Pressure 1/8 inch female NPT
    Electrical Standard banana jacks
    RTD 4 pin miniature circular
    External module 6 pin miniature circular
    Engineering Units (Configurable)
    psi, bar, mbar, kPa, kgcm2, cmH2O @ 4°C, cmH2O @ 20°C, mH2O @ 4°C, mH2O @ 20°C, inH2O @ 4°C, inH2O @ 20°C, inH2O @ 60°F, mmHg @ 0°C, inHg @ 0°C, ftH2O @ 4°C, ftH2O @ 20°C, ftH2O @ 60°F
 
     

 

  Reader Comments