Fluke 718Ex Intrinsically Safe Pressure Calibrator Fluke 718Ex Intrinsically Safe Pressure Calibrator The Fluke 718Ex is a powerful intrinsically safe self-contained portable pressure calibrator for usage in explosion endangered areas. P000214 Intrinsically Safe 0 VND Quantity: 1 Cái


  •  
  • Fluke 718Ex Intrinsically Safe Pressure Calibrator

  • Post on 23-09-2015 09:10:26 AM - 1249 Views
  • Price: Contact

  • The Fluke 718Ex is a powerful intrinsically safe self-contained portable pressure calibrator for usage in explosion endangered areas.


Quantity

Pump Up the Pressure

The Fluke 718Ex Intrinsically Safe Pressure Calibrator is a powerful intrinsically safe, self contained pressure calibrator for usage in explosion endangered areas.

 

  • Pressure Measurement Accuracy
    Fluke 718Ex 30G Range -830 mbar to 2 bar (internal sensor)
    Over pressure: 3x FS
    Resolution 0.0001 bar
    Accuracy 0.05% FS
    Functions Zero, Min, Max, Hold, Damp
    Fluke 718Ex 100G Range -830 mbar to 2 bar (internal sensor)
    Over pressure: 2x FS
    Resolution 0.001 bar
    Accuracy 0.05% FS
    Functions Zero, Min, Max, Hold, Damp
    Fluke 718Ex 300G Range -830 mbar to 20 bar (internal sensor)
    Over pressure: 375 psi, 25 bar
    Resolution 0.0001 bar
    Accuracy 0.05% FS
    Functions Zero, Min, Max, Hold, Damp
    Fluke 718Ex series Range Via pressure module connector 
    29 pressure modules, 2.5 mbar to 700 bar 
    Over pressure per pressure module specs
    Resolution Per pressure module specs
    Accuracy To 0.025% of full span, per pressure module specs
    Functions Zero, Min, Max, Hold, Damp
    Source pressure, built in pump range -820 mbar to full scale
    For use with non-corrosive gasses.
    Supported Pressure Units psi, inH2O (4°C), inH2O (20°C), kPa, cmH2O (4°C), cmH2O (20°C), bar, mbar, kg/cm², mmHg, inHg
    Media compatibility per pressure module specs
    Technical Data
    Measure accuracy Range 0-24 mA
    Resolution 0.001 mA
    Accuracy 0.02% rdg ±2 counts
    Display LCD, 5 digit pressure and current simultaneous
    Environmental Specifications
    Operating temperature -10°C to 55°C
    Storage temperature -40°C to 60°C
    Humidity (without condensation) 95% (10°C to 30°C)
    75% (30°C to 40°C)
    45% (40°C to 50°C)
    35% (50°C to 55°C)
    Operating altitude 3,000 m max
    Safety Specifications
    Safety ATEX II 1G Ex ia IIC T4 compliant
    CSA Certification I.S. Class I
    Div 1 Groups A-D , T4
    Mechanical and General Specifications
    Size (L x W x H) 216 x 94 x 66 mm (with holster)
    Weight Approx. 992 g (with holster)
    Warranty Three years, one year for pump
    Shock 1 meter drop test
    Vibration Random, 2 G, 5-500 Hz
    Battery type One 9-volt alkaline battery ANSI/NEDA 1604A or IEC 6LR619V
    Battery life 4 to 20 hours, depending on functions used
     

 